DP of Electrical Paper
Electrical paper - IEC 60450 or ASTM D4243

What solvents do I need?
The paper material is dissolved in a 50/50 mixture of water and cupriethylene diamine. Cleaning of the sample path of the automated viscometer is achieved with water followed by acetone for drying.
What is the temperature of measurement?
25 degrees C. This is achieved by using a temperature controlled bath unit. The glass viscometers for the solution measurement are suspended in the bath.
What are the installation requirements?
A laboratory environment with stable temperature between 18 and 23 degrees C. It is also important to have access to a fume hood for placing the exhaust line. The automated viscomter unit itself for DP measurement does not need to be installed in a fume hood.
How long does it take to run a DP test?
The test cycle in the automated DP measurement unit takes approximately 30 minutes. After the half hour, it is possible to load a new sample in the instrument. Before the actual measurement, the sample must be fully solubilized in the water/CED mixture. This may take up to 16 hours. The samples are typically left in a shaker overnight to be ready for measuring in the morning.
Can I measure DP without solvents?
The specific DP of electrical paper can also be determined using gel permeation chromatography or spectroscopic techniques.