PET - ASTM D4603 IV

PET - Intrinsic Viscosity - ASTM D4603

The PSL Rheotek RPV-3 determines the Inherent & intrinsic Viscosity of Poly (Ethylene Terephthalate)/PET in dilute solution using glass capillary viscometers in accordance with ASTM D4603. Meh The test method requires a 0.5% concentration of PET dissolved in 60/40 phenol/1,1,2,2-tetrachloroethane solution. Solvent and solution flow times are measured at 30 deg. C using the same 1B size of suspended-level Ubbelohde type viscometer for both measurements.

Step by step

The method requires that three successive flow times are obtained within 0.2s of the mean. The RPV-3 software automatically calculates the Intrinsic Viscosity from each single determination using the Billmeyer equation. The dual position RPV-3 (2) is an ideal configuration as well as the four position RPV-3 (4). The iSP-1 Sample Preparation system can be used in conjunction with the RPV-3 to prepare samples on a precise weight to weight basis. Dissolution is carried out using a reaction block with stirring and heating at 110 deg. C with a programmable timer. (SD-BM).

1 item(s)